ZHENGZHOU HENGTONG FURNACE CO,. LTD
  • Product categories have not been added!
1100C Quartz Tube Furnace (2" Dia.) with Temp. Calibrator and Vacuum Pump
1100C Quartz Tube Furnace (2
  • 1100C Quartz Tube Furnace (2
0

THE  vacuum tube furnace is ready to pass UL or CSA certification (UL61010 or SPE1000) on the customer side at customer cost tube furnace for heating samples up to 1100°C. The right side of the furnace system is composed of a KF25 Gas Outlet with Digital Vacuum Gauge connected to Mechanical Pump with Exhaust Filter. The 1/4'' barbed gas inlet on the left, allows the furnace to heat samples in vacuum up to 1E-2 Torr or with flowing gas. In addition, an OMEGA 1/4’’ O.D K Type Thermal Couple is included for attaching to the left flange. It can be inserted into the tube to accurately measure the sample’s temperature by the readout from the OMEGA Calibrator. The built-in precision temperature controller will provide up to 30 segments of heating and cooling steps with +/- 1 °C accuracy.

SPECIFICATIONS

Furnace Structure
  • Energy-saving high purity Al2O3 fibrous insulation liner

  • Stainless steel shells with adjustable height stand

  • Adjust three knobs on the back to flexibly change the furnace's vertical position within travel distance (from control box platform to lower end of the chamber): 4.5'' ~ 12'' (without hanger and supporter) 

  • Multiple configurations for various thermal processing needs

Power1500W
VoltageAC 110V or AC 220V 50/60 Hz  -- please select the voltage in the option bar
Working Temperature
  • Max. Temperature: 1100 °C (< 1 hr)
  • Continuous working temperature: 1000 °C

Heating RateMax. Heating Rate: 10°C /min
Heating Zone 
  • Heating Zone length: 8" (200mm) 

  • Constant Temperature Zone: 100mm (+/-1°C @ 1000 °C)

Temperature controller
  • PID automatic control and auto-tune function.

  • 30 programmable segments for precise process

  • Built-in protection for overheated and broken Thermal couple. 

  • Temperature Accuracy: +/- 1 °C

  • K-type thermal couple: K type 18" L x 1/4"Dia with Alloy sheath and Male Plug for Calibration up to 1335C

Laptop, software (Optional)
  • Brand new laptop with Microsoft Windows 10 and Microsoft Office 2013 (30 days free trial)

  • Labview Temperature Control System enables users to edit temperature profiles, manage heat-treatment recipes, and record, and plot data 

Temperature Calibrator
  • Thermal Couples: K type 18" L x 1/4" Dia with Alloy sheath and Male Plug for Calibration up to 1335°C

  • Portable Digital Temperature Calibrator with USB and RS232 Interface & Software - NIST certificate

Heating ElementsFe-Cr-Al Alloy doped by Mo
Tube Size and Materials
  • Quartz Tube with OD: 50;  ID 43:  Length: 600mm

  • Two fibrous ceramic tube blocks are included for blocking heat radiation from the inside tube. (Ceramic blocks must be fully inserted into the furnace chamber before heating)

  • Quartz blocks (optional) are ideal for cleaner and higher vacuum efficiency.

Vacuum Sealing FlangesStainless Steel Vacuum Flanges (Left with 1/4 barb fitting, Digital Vacuum Gauge and Thermocouple feedthrough, right with KF25 port and Right Angle Valve)
Fitting Ports
  • Standard 1/4‘’ Barbed fittings with needle valves.

  • For supplying high-pressure gas, you may replace the standard barbed hose fitting with the 304SS 1/4" Swagelok® Tube Fitting



Vacuum Pump and Vacuum Level
  • Double Stage Rotary Vane Vacuum Pump EQ-FYP-Pump (156L/m) with oil trap, Exhaust Filter, Bellow& KF-D25 Inlet and KF25 Clamp are included.

  • Vacuum Pressure can reach 1E-2 Torr ( 50 m-torr) 

  • 10^-5 torr can be reached by molecular pump

Dimensions10" x 15" x 16"   (26 x38 x40 cm) (D x H x W)
Net Weight100 lbs
Shipping Dimensions40"(L) x 26"(W) x 35"(H)
Warranty
  • One-year limited warranty (Consumable parts such as processing tubes and o-rings are not covered by the warranty)

Application Notes
  • Tube furnaces with quartz tubes are designed for use under vacuum and low pressure < 0.2 bars

  • Attention: A two-stage pressure regulator must be installed on the gas cylinder to limit the pressure to below 3 PSI for safe operation

  • Vacuum limit definition for all quartz tube furnaces: Vacuum pressures may only be safely used up to 1000°C

  • The flow rate for gases should be limited to<200 SCCM (or 200ml/min) for reducing thermal shocks to the tube



THE  vacuum tube furnace is ready to pass UL or CSA certification (UL61010 or SPE1000) on the customer side at customer cost tube furnace for heating samples up to 1100°C. The right side of the furnace system is composed of a KF25 Gas Outlet with Digital Vacuum Gauge connected to Mechanical Pump with Exhaust Filter. The 1/4'' barbed gas inlet on the left, allows the furnace to heat samples in vacuum up to 1E-2 Torr or with flowing gas. In addition, an OMEGA 1/4’’ O.D K Type Thermal Couple is included for attaching to the left flange. It can be inserted into the tube to accurately measure the sample’s temperature by the readout from the OMEGA Calibrator. The built-in precision temperature controller will provide up to 30 segments of heating and cooling steps with +/- 1 °C accuracy.

SPECIFICATIONS

Furnace Structure
  • Energy-saving high purity Al2O3 fibrous insulation liner

  • Stainless steel shells with adjustable height stand

  • Adjust three knobs on the back to flexibly change the furnace's vertical position within travel distance (from control box platform to lower end of the chamber): 4.5'' ~ 12'' (without hanger and supporter) 

  • Multiple configurations for various thermal processing needs

Power1500W
VoltageAC 110V or AC 220V 50/60 Hz  -- please select the voltage in the option bar
Working Temperature
  • Max. Temperature: 1100 °C (< 1 hr)
  • Continuous working temperature: 1000 °C

Heating RateMax. Heating Rate: 10°C /min
Heating Zone 
  • Heating Zone length: 8" (200mm) 

  • Constant Temperature Zone: 100mm (+/-1°C @ 1000 °C)

Temperature controller
  • PID automatic control and auto-tune function.

  • 30 programmable segments for precise process

  • Built-in protection for overheated and broken Thermal couple. 

  • Temperature Accuracy: +/- 1 °C

  • K-type thermal couple: K type 18" L x 1/4"Dia with Alloy sheath and Male Plug for Calibration up to 1335C

Laptop, software (Optional)
  • Brand new laptop with Microsoft Windows 10 and Microsoft Office 2013 (30 days free trial)

  • Labview Temperature Control System enables users to edit temperature profiles, manage heat-treatment recipes, and record, and plot data 

Temperature Calibrator
  • Thermal Couples: K type 18" L x 1/4" Dia with Alloy sheath and Male Plug for Calibration up to 1335°C

  • Portable Digital Temperature Calibrator with USB and RS232 Interface & Software - NIST certificate

Heating ElementsFe-Cr-Al Alloy doped by Mo
Tube Size and Materials
  • Quartz Tube with OD: 50;  ID 43:  Length: 600mm

  • Two fibrous ceramic tube blocks are included for blocking heat radiation from the inside tube. (Ceramic blocks must be fully inserted into the furnace chamber before heating)

  • Quartz blocks (optional) are ideal for cleaner and higher vacuum efficiency.

Vacuum Sealing FlangesStainless Steel Vacuum Flanges (Left with 1/4 barb fitting, Digital Vacuum Gauge and Thermocouple feedthrough, right with KF25 port and Right Angle Valve)
Fitting Ports
  • Standard 1/4‘’ Barbed fittings with needle valves.

  • For supplying high-pressure gas, you may replace the standard barbed hose fitting with the 304SS 1/4" Swagelok® Tube Fitting



Vacuum Pump and Vacuum Level
  • Double Stage Rotary Vane Vacuum Pump EQ-FYP-Pump (156L/m) with oil trap, Exhaust Filter, Bellow& KF-D25 Inlet and KF25 Clamp are included.

  • Vacuum Pressure can reach 1E-2 Torr ( 50 m-torr) 

  • 10^-5 torr can be reached by molecular pump

Dimensions10" x 15" x 16"   (26 x38 x40 cm) (D x H x W)
Net Weight100 lbs
Shipping Dimensions40"(L) x 26"(W) x 35"(H)
Warranty
  • One-year limited warranty (Consumable parts such as processing tubes and o-rings are not covered by the warranty)

Application Notes
  • Tube furnaces with quartz tubes are designed for use under vacuum and low pressure < 0.2 bars

  • Attention: A two-stage pressure regulator must be installed on the gas cylinder to limit the pressure to below 3 PSI for safe operation

  • Vacuum limit definition for all quartz tube furnaces: Vacuum pressures may only be safely used up to 1000°C

  • The flow rate for gases should be limited to<200 SCCM (or 200ml/min) for reducing thermal shocks to the tube