HTTF-1200-S Sliding Tube Furnace is a single-temperature zone rapid thermal processing electric sliding furnace. This electric furnace has a maximum heating rate of 10℃/min and is equipped with a sliding rail at the bottom, which allows the furnace to slide from one side of the tube to the other for rapid heating and cooling. This RTP furnace can also be used for the CVD growth of graphene and carbon nanotubes.

HTTF-1200-S Sliding Tube Furnace is a single-temperature zone rapid thermal processing electric sliding furnace. This electric furnace has a maximum heating rate of 10/min and is equipped with a sliding rail at the bottom, which allows the furnace to slide from one side of the tube to the other for rapid heating and cooling. This RTP furnace can also be used for the CVD growth of graphene and carbon nanotubes. It uses a touch screen and PLC control, which allows for adjustable speed.

 

Power

AC 208-240V Single Phase, 50/60 Hz

3 kW

Heating

Maximum:1200< 0.5 hrs

Continuous Working: 1100 or less

Recommended heating rate ≤10/min

Heating zone length: 400mm

Thermocouple

 K-type

Specs

Slide distance: 400mm (can be customized at extra cost)

Adjustable speed range: 10-100mm/s (can be customized at extra cost) and can also be set to move back and forth

The heating will automatically slide to the far right from the far left according to the temperature program after reaching a certain temperature

The maximum heating rate for the left sample area: 3/s

Rapid cooling rate for samples inside the furnace: 3/s (500-100), 0.7/s (500-200)

Temperature Controller

Two PID automatic controllers with 50 programmable segments for control of heating & cooling rate, and dwell time.

Built-in overheating & thermocouple broken protection.

+/- 1 ºC temperature accuracy.

DB9 PC Communications Port.

PC Control:  

A laptop PC can be ordered for $500 together with the furnace. We recommend ordering a computer (laptop) from us to avoid software installation problems as we will perform installation and testing before shipping.

Vacuum Flanges

One set of stainless steel sealed flanges

The flanges and quartz tube are sealed with O-rings, pressure rings, and O-rings, and tightened by threaded compression to ensure sealing

The left flange has a KF25 vacuum interface, needle valve, and barb venting port

The right flange has a needle valve, G1/4 air inlet, and mechanical pressure gauge.

Quartz Tube Options

25 O.D x 19 I.D x 1400 L (mm)

50 O.D x 44 I.D x 1400 L (mm)

60 O.D x 55 I.D x 1400 L (mm)

80 O.D x 75 I.D x 1400 L (mm)

100 O.D x 95 I.D x 1400 L (mm)

Vacuum Level 

 Limited by a vacuum pump

10^-2 torr can be achieved by a mechanical pump

10^-5 torr can be reached by a molecular pump

 The vacuum pump is not included, please order separately 

Dimensions

1900 * 500 * 680mm

Net Weight

~ 210 Kg

Warranty

One-year limited manufacturer's warranty (Consumable parts such as processing tubes, o-rings and heating elements are not covered by the warranty, please order replacements at related products below).

Application Notes

Tube furnaces with quartz tubes are designed for use under vacuum and low pressure < 0.12 MPa 

Attention: A two-stage pressure regulator must be installed on the gas cylinder to limit the pressure to below 3 PSI for safe operation.

 Vacuum pressures may only be safely used up to 1000°C

 


HTTF-1200-S Sliding Tube Furnace is a single-temperature zone rapid thermal processing electric sliding furnace. This electric furnace has a maximum heating rate of 10/min and is equipped with a sliding rail at the bottom, which allows the furnace to slide from one side of the tube to the other for rapid heating and cooling. This RTP furnace can also be used for the CVD growth of graphene and carbon nanotubes. It uses a touch screen and PLC control, which allows for adjustable speed.

 

Power

AC 208-240V Single Phase, 50/60 Hz

3 kW

Heating

Maximum:1200< 0.5 hrs

Continuous Working: 1100 or less

Recommended heating rate ≤10/min

Heating zone length: 400mm

Thermocouple

 K-type

Specs

Slide distance: 400mm (can be customized at extra cost)

Adjustable speed range: 10-100mm/s (can be customized at extra cost) and can also be set to move back and forth

The heating will automatically slide to the far right from the far left according to the temperature program after reaching a certain temperature

The maximum heating rate for the left sample area: 3/s

Rapid cooling rate for samples inside the furnace: 3/s (500-100), 0.7/s (500-200)

Temperature Controller

Two PID automatic controllers with 50 programmable segments for control of heating & cooling rate, and dwell time.

Built-in overheating & thermocouple broken protection.

+/- 1 ºC temperature accuracy.

DB9 PC Communications Port.

PC Control:  

A laptop PC can be ordered for $500 together with the furnace. We recommend ordering a computer (laptop) from us to avoid software installation problems as we will perform installation and testing before shipping.

Vacuum Flanges

One set of stainless steel sealed flanges

The flanges and quartz tube are sealed with O-rings, pressure rings, and O-rings, and tightened by threaded compression to ensure sealing

The left flange has a KF25 vacuum interface, needle valve, and barb venting port

The right flange has a needle valve, G1/4 air inlet, and mechanical pressure gauge.

Quartz Tube Options

25 O.D x 19 I.D x 1400 L (mm)

50 O.D x 44 I.D x 1400 L (mm)

60 O.D x 55 I.D x 1400 L (mm)

80 O.D x 75 I.D x 1400 L (mm)

100 O.D x 95 I.D x 1400 L (mm)

Vacuum Level 

 Limited by a vacuum pump

10^-2 torr can be achieved by a mechanical pump

10^-5 torr can be reached by a molecular pump

 The vacuum pump is not included, please order separately 

Dimensions

1900 * 500 * 680mm

Net Weight

~ 210 Kg

Warranty

One-year limited manufacturer's warranty (Consumable parts such as processing tubes, o-rings and heating elements are not covered by the warranty, please order replacements at related products below).

Application Notes

Tube furnaces with quartz tubes are designed for use under vacuum and low pressure < 0.12 MPa 

Attention: A two-stage pressure regulator must be installed on the gas cylinder to limit the pressure to below 3 PSI for safe operation.

 Vacuum pressures may only be safely used up to 1000°C